HT2800T_en_c.doc Page 237 of 275
below ), must be extracted from the syringe barrel (number 8 in figure below ). Pull
them out from the needle side and do it very gently without forcing. Please note that
all of the elements should be separated naturally during the syringe barrel extraction
from the heating body.
Figure 192: Syringe warmer assembly disassembling_2
5) Replace the syringe barrel with a new one.
6) Re-assemble the syringe warmer assembly. Start by placing the lower spacer (number
9 in figure below) inside the heating body (number 4 in figure below). This operation
must not be carried out for 5ml syringe.
Figure 193: Lower spacers positioning
7) Also place the insulating gasket (number 5 in figure below) inside the heating
body (number 4 in figure below).