EasyManua.ls Logo

iTero Element 2 - Page 125

iTero Element 2
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© 2019 Align Technology, Inc. All rights reserved.
103
iTero Element 2 and iTero Element Flex User Manual
Test
Standard
Severity level/lines
Test
results
Emission (per ETSI EN 301 489-1)
Relevant for iTero Element 2 only
Conducted emissions on
mains terminals in freq.
range 150 kHz - 30 MHz
ETSI EN 301 489-
1;
EN 55032
Group 1 Class B
230 VAC mains
Complies
Radiated emissions
in freq. range 30 - 6000 MHz
ETSI EN 301 489-
1;
EN 55032
Class B
Complies
Harmonic current test
ETSI EN 301 489-
1;
EN 61000-3-2
230 VAC mains
Complies
Flicker tests
ETSI EN 301 489-
1
EN 61000-3-3
230 VAC mains
Complies
Immunity (per ETSI EN 301 489-1)
Relevant for iTero Element 2 only
Immunity from Electrostatic
discharge (ESD)
EN 61000-4-2
4 kV contact discharge
8 kV air discharge
Complies
Immunity from radiated
electromagnetic fields
EN 61000-4-3
3.0 V/m,
80 MHz 6.0 GHz,
80% AM, 1 kHz
Complies
Immunity from Electrical
Fast transient (EFT)
EN 61000-4-4
AC mains: ± 1.0 kV;
Tr/Th 5/50 ns, 5 kHz
Complies
Immunity from Surge
EN 61000-4-5
AC mains: ± 1.0 kV DM / ± 2.0 kV CM,
Tr/Th 1.2/50 (8/20) s
Complies
Immunity from conducted
disturbances induced by
radio-frequency fields
EN 61000-4-6
AC mains: 3.0 VRMS;
0.15÷80 MHz, 80% AM @ 1 kHz
Complies
Immunity from Voltage
interruptions
EN 61000-4-11
AC mains:
0 % - 0.5 cycle & 1 cycle;
70% - 25 cycles; 0% - 250 cycles
Complies