EasyManua.ls Logo

Keithley 2010

Keithley 2010
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[:SENSe[1]]
:TEMPerature
:FRTD
:TYPE <name>
:TYPE?
:RZERo <NRf>
:RZERo?
:ALPHa <NRf>
:ALPHa?
:BETA <NRf>
:BETA?
:DELTa <NRf>
:DELTa?
Path to configure FRTD sensor.
Select FRTD sensor type (PT100, D100, F100, PT3916,
PT385, USER)
Query FRTD sensor type.
Specify RZERo value (0 to 10,000).
Query RZERo value.
Specify ALPHa value (0 to 0.01).
Query ALPHa value.
Specify BETA value (0 to 1.00).
Query BETA value.
Specify DELTa value (0 to 5.00).
Query DELTa value.
PT100
100.00
0.00385
0.11100
1.50700
:FREQuency
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
Path to configure frequency.
Path to select the threshold voltage range:
Select threshold range (0 to 1010).
Query threshold range.
Specify reference (0 to 1.5e7).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
10
0
OFF
7
:PERiod
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
Path to configure period.
Path to select the threshold voltage range:
Select threshold range (0 to 1010).
Query threshold range.
Specify reference (0 to 1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
10
0
OFF
7
:DIODe
:CURRent
:RANGe
[:UPPer] <NRf>
[:UPPer]?
Paths to configure diode test:
Path to select range.
Select range (0 to 1e-3).
Query range.
1e-3
Table 5-6 (cont.)
SENSe command summary
Command Description
Default
parameter
SCPI
5-16 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2010