5-62 SCPI Command Reference
:ENABle <list>
:STATus:QUEue:ENABle <list> Enable messages for Error Queue
Parameter <list> = (numlist)
where numlist is a specified list of messages that you wish to enable for the
Error Queue.
Query :ENABle? Query list of enabled messages
Description On power-up, all error messages are enabled and will go into the Error Queue
as they occur. Status messages are not enabled and will not go into the queue.
This command is used to specify which messages you want enabled.
Messages not specified will be disabled and prevented from entering the
queue.
When this command is sent, all messages will first be disabled, then the
messages specified in the list will be enabled. Thus, the returned list
(:ENABle?) will contain all the enabled messages.
Messages are specified by numbers (see Appendix B). The following
examples show various forms for expressing a message numlist.
Numlist = -110 Single message.
-110, -140, -222 Messages separated by commas.
-110:-222 Range of messages (-110 through -222).
-110:-222, -230 Range entry and single entry separated by a
comma.
NOTE To disable all messages from entering the Error Queue, send the
following command:
:stat:que:enab()

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz