Model 2700 Multimeter/Switch System User’s Manual Basic DMM Operation 3-19
6. Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed (or press AUTO for
autoranging). For manual ranging, use the lowest possible range for the best
resolution.
7. To measure another amps channel, repeat steps 5 and 6.
8. When finished, press OPEN if there is a channel closed.
NOTE When you have an amps-only channel closed, you cannot select a non-amps
function. For example, if channel 21 of the Model 7700 is closed, you cannot
select the DCV function (“INVALID FUNC” displayed).
AMPS fuse replacement (front panel AMPS input)
WARNING Make sure the instrument is disconnected from the power line and
other equipment before replacing the AMPS fuse.
1. Turn off the power and disconnect the power line and test leads.
2. From the front panel, gently push in the AMPS fuse holder with a flat blade
screwdriver and rotate the fuse holder one-quarter turn counterclockwise.
3. Remove the fuse and replace it with the same type (3A, 250V, fast-blow, 5 ×
20mm). The Keithley part number is FU-99-1.
CAUTION Do not use a fuse with a higher current rating than specified or
instrument damage may occur. If the instrument repeatedly blows
fuses, locate and correct the cause of the trouble before replacing the
fuse.
4. Install the new fuse by reversing the procedure above.
NOTE For the Model 7700 switching module and other similar modules that support
the amps function, there are solder mount amps fuses. See the Model 2700
Service Manual for fuse replacement information.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD