Model 2700 Multimeter/Switch System User’s Manual Scanning 7-33
Scanning examples
The following scanning examples assume that the Model 7700 switching module is
installed in slot 1 of the mainframe.
Tables are used for the procedure steps to configure and run scan examples. The left side
of the table provides the front panel procedure, while the right side shows the equivalent
remote programming commands. Where appropriate, menu sequences are provided to
summarize a front panel operation or selection. For example:
SHIFT SETUP > RESTORE: FACT
For the above menu sequence, press SHIFT and then SETUP to access the menu, use the
edit keys (,
, Δ, and ) to display RESTORE: FACT, and then press ENTER to
select it.
External trigger scan
For this example, an external instrument is used to trigger the start of the 2-channel scan.
Trigger pulse requirements and trigger cable connections are covered in
Section 7.
NOTE For this example, the front panel TRIG key can be used in place of an external
input trigger. Each time the TRIG key is pressed, the 2-channel scan will run.
One channel (101) measures temperature and the other channel (102) measures resistance.
The two readings are stored in the buffer. Each time the scan is run, the two readings will
be appended (added) to the buffer.
A type K thermocouple is used to measure temperature. Since the internal cold (reference)
junction of the Model 7700 is being used, the thermocouple can be connected directly to
the screw terminals of the switching module.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals