3-40 Basic DMM Operation Model 2700 Multimeter/Switch System User’s Manual
Temperature measurement configuration
The Model 2700 is configured to measure temperature from the temperature measurement
configuration menu. Use the following general rules to navigate through the menu
structure:
Press SHIFT and then SENSOR to enter the menu structure.
Cursor position is indicated by a flashing menu item or parameter. Cursor position
is controlled by the and
keys.
With the cursor on a menu item or parameter, use the Δ and keys to scroll
through the available options.
A displayed menu item and parameter is selected by pressing ENTER.
You can exit from the menu structure by pressing EXIT. However, any ENTERed
selections will apply.
Thermocouple temperature measurement configuration
The steps to configure thermocouple measurements are provided in Table 3-3. After
pressing SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
NOTE An INT card is a switching module that has an internal reference junction (i.e.,
Model 7700). The INT reference junction setting cannot be selected if there is
not at least one INT card installed in the unit. With no INT cards installed,
selecting INT will cause the “NO INT CARDS” message to be displayed briefly.
With at least one INT card installed, the INT reference junction can be selected.
However, if you select it for the front panel inputs, or for a switching module that
does not have an internal reference junction (i.e., Model 7702), the simulated
(SIM) reference junction will instead be used and the “ERR” annunciator will
turn on.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals