F-8 Temperature Equations Model 2700 Multimeter/Switch System User’s Manual
RTD equations
The temperature vs. resistance readings listed in the RTD reference tables are calculated
using the Callendar-Van Dusen equation. There are two equations based on different
temperature ranges. There is an equation for the -200° to 0°C range and one for the 0° to
630°C range.
Equation for -200° to 0°C temperature range
R
RTD
= R
0
[1 + AT + BT
2
+ CT
3
(T-100)]
where: R
RTD
is the calculated resistance of the RTD
R
0
is the known RTD resistance at 0°C
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = -1 (alpha)(delta)(1e-4)
C = -1 (alpha)(beta)(1e-8)
The alpha, beta, and delta values are listed in Table F-10.
Equation for 0° to 630°C temperature range
R
RTD
= R
0
(1 + AT + BT
2
)
where: R
RTD
is the calculated resistance of the RTD
R
0
is the known RTD resistance at 0°C
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = -1 (alpha)(delta)(1e-4)
The alpha and delta values are listed in Table F-10.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals