4-20 Range, Digits, Rate, Bandwidth, and Filter Model 2700 Multimeter/Switch System User’s Manual
Remote programming — filter
Filter commands
The filter commands are listed in Table 4-6. Additional information on these commands
follow the table.
NOTE Query commands are not included in Table 4-6. All commands for the SENSe
subsystem are provided in Table 15-5.
Table 4-6
Filter commands
Commands
1, 4
Description
5
Default
DCV filter commands
[SENSe[1]] Optional root command.
:VOLTage[:DC]:AVERage:TCONtrol <name> Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:VOLTage[:DC]:AVERage:WINDow <NRf> Set filter window in %; <NRf> = 0 to 10. 0.1
:VOLTage[:DC]:AVERage:COUNt <n> [, clist] Specify filter count; <n> = 1 to 100. 10
:VOLTage[:DC]:AVERage:STATe <b> [, clist] Enable or disable the filter. (Note 3)
ACV filter commands
[SENSe[1]] Optional root command.
:VOLTage:AC:AVERage:TCONtrol <name> Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:VOLTage:AC:AVERage:WINDow <NRf> Set filter window in %; <NRf> = 0 to 10. 0.1
:VOLTage:AC:AVERage:COUNt <n> [, clist] Specify filter count; <n> = 1 to 100. 10
:VOLTage:AC:AVERage:STATe <b> [, clist] Enable or disable the filter. (Note 3)
DCI filter commands
[SENSe[1]] Optional root command.
:CURRent[:DC]:AVERage:TCONtrol <name> Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:CURRent[:DC]:AVERage:WINDow <NRf> Set filter window in %; <NRf> = 0 to 10. 0.1
:CURRent[:DC]:AVERage:COUNt <n>
[, clist]
Specify filter count; <n> = 1 to 100. 10
:CURRent[:DC]:AVERage:STATe <b> [, clist] Enable or disable the filter. (Note 3)
ACI filter commands
[SENSe[1]] Optional root command.
:CURRent:AC:AVERage:TCONtrol <name> Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:CURRent:AC:AVERage:WINDow <NRf> Set filter window in %; <NRf> = 0 to 10. 0.1
:CURRent:AC:AVERage:COUNt <n> [, clist] Specify filter count; <n> = 1 to 100. 10
:CURRent:AC:AVERage:STATe <b> [, clist] Enable or disable the filter. (Note 3)

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD