13-4 SCPI Signal Oriented Commands Model 2700 Multimeter/Switch System User’s Manual
CONFigure:<function> [<rang>], [<res>], [<clist>]
CONFigure:VOLTage[:DC] [<rang>], [<res>], [<clist>] Configure DCV
CONFigure:VOLTage:AC [<rang>], [<res>], [<clist>] Configure ACV
CONFigure:CURRent[:DC] [<rang>], [<res>], [<clist>] Configure DCI
CONFigure:CURRent:AC [<rang>], [<res>], [<clist>] Configure ACI
CONFigure:RESistance [<rang>], [<res>], [<clist>] Configure Ω2
CONFigure:FRESistance [<rang>], [<res>], [<clist>] Configure Ω4
CONFigure:FREQuency [<rang>], [<res>], [<clist>] Configure FREQ
CONFigure:PERiod [<rang>], [<res>], [<clist>] Configure PERIOD
CONFigure:TEMPerature [<clist>] Configure TEMP
CONFigure:CONTinuity [<clist>] Configure CONT
Parameters <rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See
the “NOTES” that follow Table 13-1 for additional information.
<res> = 0.1 i.e., 100.0 V (3H digits)
0.01 i.e., 10.00 V (3H digits)
0.001 i.e., 1.000 V (3H digits)
0.0001 i.e., 1.0000 V (4H digits)
0.00001 i.e., 1.00000 V (5H digits)
0.000001 i.e., 1.000000 V (6H digits)
The resolution of the <res> parameter value and the selected range
sets the number of display digits. As shown above, with the 100V
range selected and <res> = 0.1, a 100V reading will be displayed
as 100.0 V (3H digits).
The display will default to 3H digits when using parameter values
that attempt to set the display below 3H digits. For example, a 10V
reading using <res> = 0.1 for the 10V range is displayed as 10.00
V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7H digits is ignored, and generates error -221
(settings conflict error).
The <res> parameter is ignored when a <clist> is included in
the command string. Resolution for the scanlist channel(s) is
determined by the present setting for the specified function and
by the present resolution setting for the specified function.
See the “NOTES” that follow Table 13-1 for additional
information.
<clist> = Channel(s) in the scanlist to be configured. When the channel
list parameter (<clist>) is included, the present instrument
settings are not affected. Instead, the channel(s) in the <clist>
for the specified function is configured. See the “NOTES” that
follow
Table 13-1 for additional information.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD