Model 2700 Multimeter/Switch System User’s Manual Triggering 8-5
If MANual is chosen, also enter the duration of the delay in the hour/minute/second
format using the ,
, Δ, and keys. The maximum is 99H:99M:99.999S:. Note that
pressing the AUTO key sets the delay to 0.001 sec. Press ENTER to accept the delay or
EXIT for no change.
Device action
The primary device action is a measurement. However, the device action block could
include the following additional actions (
Figure 8-2):
Figure 8-2
Device action
Filtering — If the repeating filter is enabled, the instrument samples the specified
number of reading conversions to yield single, filtered reading. Only one reading
conversion is performed if the filter is disabled, or after the specified number of
reading conversions for a moving average filter is reached. After a reading (Rdg) is
procured, operation proceeds to Hold.
Hold — The Hold feature is used to screen out reading anomalies. When enabled,
the user selects a window and count for Hold. In general, when a reading is outside
the window it is rejected, operation loops back to the beginning of the Device
Action as shown in
Figure 8-2. The hold count specifies how many readings have
to be within the window before it is accepted. See “Reading hold (autosettle),
page 8-6, for operation details. After a Hold Reading is acquired, operation
proceeds to Channel Closure.
Channel Closure — When scanning, the last device action is channel control.
Using the hold feature provides an auto settling time for switching relays. Each
open/close transition will restart the hold process and a reading for each channel
will not occur until the relay settles.
Rdg Hold Chan
Filter
DEVICE ACTION
To Output Trigger
Block of Figure 8-1
From Delay Block
of Figure 8-1

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals