B-6 Model 7700 Connection Guide Model 2700 Multimeter/Switch System User’s Manual
Figure B-3
Model 7700 screw terminal channel designations
Wiring procedure
Use the following procedure to wire the Model 7700 module. Make all connections using
correct wire size (up to 20 AWG). Also, make sure to add supplementary insulation
around the harness for voltages above 42V peak (
Figure B-4).
WARNING All wiring and supplementary insulation must be rated for the
maximum voltage in the system. For example, if 1000V is applied to
the front terminals of the DMM, the plug-in module wiring must be
rated for 1000V.
1. Make sure all power is discharged from the Model 7700 module.
2. Access the screw terminals (Figure B-2).

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD