Model 2700 Multimeter/Switch System User’s Manual Basic DMM Operation 3-25
Performing offset-compensated ohms measurements
Offset-compensated ohms can only be performed on the Ω4 function using the 100Ω, 1kΩ,
or 10k
Ω range. Make sure you use 4-wire connections to the DUT (see “Connections,
page 3-8).
NOTE Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1. If a switching channel is presently closed (displayed), press OPEN to open it.
2. Select the 4-wire ohms measurement function by pressing Ω4, and enable offset
compensated ohms by pressing SHIFT and then OCOMP (OCOMP annunciator
turns on).
3. Use the RANGE up and down keys to select the 100Ω, 1kΩ, or 10kΩ range, or press
AUTO to enable auto range. If using auto range, offset compensated ohms mea
-
surements will not be performed if the instrument goes to the 100kΩ or higher
range.
4. Perform steps 4 through 8 of the “Standard resistance measurements,” page 3-23,
procedure.
NOTE The OCOMP annunciator will flash when the instrument is on an invalid range
(100k
Ω through 100MΩ ranges) for offset-compensated ohms. Normal ohms
measurements will instead be performed.
For buffer recall, there is no way to distinguish between a normal ohms reading
and an offset-compensated ohms reading. The OCOMP annunciator (off, on, or
flashing) has no significance for recalled resistance readings that are displayed.
Buffer operation is covered in Section 6.
With offset-compensated ohms enabled, it will be “remembered” by the Ω4 func-
tion after you change measurement functions (i.e., DCV). When Ω4 is again
selected, offset-compensated ohms will be enabled.
Measurement methods
The Model 2700 uses two methods to measure resistance:
Constant-current source method (100Ω through 1MΩ ranges) – Sources a
constant-current to the DUT. Voltage is measured by the Model 2700 and
resistance is then calculated (R = V/I).
Ratiometric method (10MΩ and 100MΩ ranges) – Test current is generated by a
0.7µA source in parallel with a 10MΩ reference resistor.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD