9-4 Limits and Digital I/O Model 2700 Multimeter/Switch System User’s Manual
Scanning
When a simple scan is configured, the present limit values and state will apply to all
channels in the scan. When an advanced scan is configured, each channel can have its own
unique limits configuration. Details to configure and run a scan are provided in
Section 7.
For remote programming, the <clist> parameter is used to configure channels for a scan.
Basic limits operation
The limits configuration is the same for all functions. For example, if a reading limit is set
to 1, that will equate to 1V for a voltage function, 1A for a current function, and 1
Ω for an
ohms function.
Setting limits
1. Press SHIFT and then LIMITS to display the high limit for Limit 1 (HI1).
2. Use , , Δ, and to key in the HI1 limit and press ENTER.
When editing a reading, use the range designator (m, ^, K, M, or G) as a multiplier.
With the cursor on the range designator, each press of
Δ or will increase or
decrease the reading by a factor of 10.
3. Key in the low limit for Limit 1 (LO1) and press ENTER.
4. Key in the high limit for Limit 2 (HI2) and press ENTER.
5. Key in the low limit for Limit 2 (LO2) and press ENTER. The instrument will
return the normal measurement state.
Beeper settings
The beeper is configured from the OUTPUT menu (shown in Table 9-1) as follows:
1. Press SHIFT and then OUTPUT.
2. Use the Δ or key to display the present beeper (BEEP) setting; NEVER,
INSIDE, or OUTSIDE.
3. Press to position the cursor on the present beeper setting, use the Δ or key to
display the desired setting, and press ENTER. The instrument will return to the
normal measurement state.
NOTE Remote programming cannot be used to set the beeper. It can only be set from
the front panel.
Enabling/disabling limits
Press SHIFT and then ON/OFF to display the present state (off or on) of limits. To enable
limits, use the
Δ or key to display “LIMITS: ON” and press ENTER.
To disable limits, again press SHIFT and then ON/OFF, select “LIMITS: OFF” and press
ENTER.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals