2-4 Close/Open Switching Module Channels Model 2700 Multimeter/Switch System User’s Manual
Connections
WARNING Connection information for switching modules is intended for quali-
fied service personnel. Do not attempt to connect DUT or external cir-
cuitry to a switching module unless qualified to do so.
WARNING To prevent electric shock that could result in serious injury or death,
adhere to following safety precautions:
Before making or breaking connections to the switching module,
make sure the Model 2700 is turned off and power is removed from
all external circuitry.
Do not connect signals that will exceed the maximum
specifications of switching module. Specifications for the
Model
7700 are provided in Appendix A.
WARNING If both the front panel terminals and the switching module terminals
are connected at the same time, the test lead insulation must be rated
to the highest voltage that is connected. For example, if 1000V is
connected to the front panel input, the test lead insulation for the
switching module must also be rated for 1000V.
WARNING Dangerous arcs of an explosive nature in a high energy circuit can
cause severe personal injury or death. If the multimeter is connected to
a high energy circuit when set to a current range, low resistance range,
or any other low impedance range, the circuit is virtually shorted.
Dangerous arcing can result even when the multimeter is set to a
voltage range if the minimum voltage spacing is reduced in the
external connections. For details to safely make high energy
measurements, see
Section 3, ““High energy circuit safety
precautions,” page 3-3.”
As described in the International Electrotechnical Commission (IEC)
Standard IEC 664, the Model 2700 is Installation Category I and must
not be connected to mains.
For the Model 7700, detailed connection and wiring information is provided in
Appendix B of this manual (Model 7700 Connection Guide).

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD