EasyManua.ls Logo

Keithley 2700 - Optimizing Measurement Speed; DC Voltage, DC Current, and Resistance; AC Voltage and AC Current; Temperature

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-10 Specifications Model 2700 Multimeter/Switch System User’s Manual
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups
restored.
DC voltage, DC current, and resistance:
Select 3H digits, 0.01 PLC, filter OFF, fixed range.
AC voltage and AC current:
Select 3H digits, 0.01 PLC, filter OFF, fixed range.
Temperature:
Select 3H digits, 0.01 PLC, filter OFF.
For all functions, turn off the display and autozero, and set the trigger delay to zero. Use
the :SAMPle:COUNt and READ? bus commands.