13-8 SCPI Signal Oriented Commands Model 2700 Multimeter/Switch System User’s Manual
MEASure:<function>? [<rang>], [<res>], [<clist>]
MEASure:VOLTage[:DC]? [<rang>], [<res>], [<clist>] Measure DCV
MEASure:VOLTage:AC? [<rang>], [<res>], [<clist>] Measure ACV
MEASure:CURRent[:DC]? [<rang>], [<res>], [<clist>] Measure DCI
MEASure:CURRent:AC? [<rang>], [<res>], [<clist>] Measure ACI
MEASure:RESistance? [<rang>], [<res>], [<clist>] Measure Ω2
MEASure:FRESistance? [<rang>], [<res>], [<clist>] Measure Ω4
MEASure:FREQuency? [<rang>], [<res>], [<clist>] Measure FREQ
MEASure:PERiod? [<rang>], [<res>], [<clist>] Measure PERIOD
MEASure:TEMPerature? [<clist>] Measure TEMP
MEASure:CONTinuity? [<clist>] Measure CONT
Parameters <rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See the
“NOTES” that follow Table 13-1 for additional information.
<res> = 0.1 i.e., 100.0 V (3H digits)
0.01 i.e., 10.00 V (3H digits)
0.001 i.e., 1.000 V (3H digits)
0.0001 i.e., 1.0000 V (4H digits)
0.00001 i.e., 1.00000 V (5H digits)
0.000001 i.e., 1.000000 V (6H digits)
The resolution of the <res> parameter value and the selected
range sets the number of display digits. As shown above, with
the 100V range selected and <res> = 0.1, a 100V reading will
be displayed as 100.0 V (3H digits).
The display will default to 3H digits when using parameter
values that attempt to set the display below 3H digits. For
example, a 10V reading using <res> = 0.1 for the 10V range is
displayed as 10.00 V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7H digits is ignored, and generates error -221
(settings conflict error).
See the “NOTES” that follow Table 13-1 for additional
information.
<clist> = Single channel only. When included, this is the channel to be
closed and measured.
Description The MEASure? command combines all of the other signal oriented
measurement commands to perform a “one-shot” measurement and
acquire the reading. If the <clist> parameter is included, the specified
channel will close before performing the measurement.
When a MEASure? command is sent, the specified function is selected.
If specified, range and resolution will also set.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD