Model 2700 Multimeter/Switch System User’s Manual SCPI Signal Oriented Commands 13-9
Depending on the specified resolution, the measurement rate is set as
follows:
6H-digits NPLC = 1.0 Medium
5H-digits NPLC = 0.1 Fast
3H or 4H-digits NPLC = 0.01 >Fast
If resolution is not specified, 6H-digit resolution and medium speed will
be selected when MEAS? is sent.
All other instrument settings related to the selected function are reset to
the *RST defaults.
If only MEASure? is sent, the Medium measurement rate is selected.
NOTE If a function is not specified, the command executes as if the present function is
specified. For example, assume the Ω2 function is presently selected. When
MEAS? is sent, the instrument resets to the *RST defaults for the Ω2 function,
and then performs a measurement.
When this command is sent, the following commands execute in the
order that they are presented:
ABORt
CONFigure:<function>
READ?
When ABORt is executed, the instrument goes into the idle state if
continuous initiation is disabled. If continuous initiation is enabled, the
operation re-starts at the beginning of the Trigger Model.
When CONfigure is executed, the MEASure? parameters (<rang>,
<res> and <clist>) are executed and the instrument goes into the “one-
shot” measurement mode. It is similar to sending the CONFigure
command with no <clist> parameter. See CONFigure for more details.
When READ? is executed, its operations will then be performed. In
general, another ABORt is performed, then an INITiate, and finally a
FETCh? to acquire the reading. See READ? for more details.
Programming examples:
Programming example #1 — The following command measures DCV
on channel 101 using the 10V range with 3H digit display resolution:
MEAS:VOLT? 10, 0.01, (@101)
Programming example #2 — The following command measures DCV
on the 100V range:
MEAS:VOLT? 100