7-32 Scanning Model 2700 Multimeter/Switch System User’s Manual
e. SAMPle:COUNt and TRIGger:COUNt — Sample count specifies the number of
readings to scan and store in the buffer, while the trigger count specifies the
number of scans to perform.
If the sample count is greater than the number of channels in the scan list (scan list
length), operation wraps around to the beginning of the scan list and continues. For
example, assume the scan list is made up of channels 101, 102, and 103, and the
sample count is set to 4. After channels 101, 102, and 103 are scanned, operation
loops around to scan channel 101 again. The first and last readings in the buffer
will be channel 101.
When performing multiple scans (trigger count >1), sample readings overwrite the
readings stored for the previous scan.
Continuous initiation must be disabled in order to set the sample counter >1 (see
Reference c).
f. INITiate:CONTinuous, INITiate and READ? — In order to initiate a single scan
cycle using INITiate or READ?, continuous initiation must be disabled. If you send
INIT or READ? with continuous initiation enabled, error -213 (Init ignored) will
occur.
You cannot use READ? or INITiate if sample count >1, AND there are readings
stored in the buffer by the TRAC command, or by front panel data store operation
(error -225, out of memory). Either set the sample count to one or clear the buffer
(TRACe:CLEar).
Scanning programming example
NOTE The following example can be run from the KE2700 Instrument Driver using the
example named “ScanChan” in
Table H-1 of Appendix H.
The following program will scan 10 channels (101 through 110):
TRAC:CLE ' Clear buffer.
INIT:CONT OFF ' Disable continuous initiation.
TRIG:SOUR IMM ' Select the immediate control source.
TRIG:COUN 1 ' Set to perform one scan.
SAMP:COUN 10 ' Set to scan 10 channels.
ROUT:SCAN (@101:110) ' Set scan list channels; 101 through 110.
ROUT:SCAN:TSO IMM ' Start scan when enabled and triggered.
ROUT:SCAN:LSEL INT ' Enable scan.
READ? ' Trigger scan and request the readings.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals