9-14 Limits and Digital I/O Model 2700 Multimeter/Switch System User’s Manual
b. CALCulate3:LIMit1:FAIL?
CALCulate3:LIMit2:FAIL?
These commands are used to query the results of Limit 1 and Limit 2:
0 = Passing (reading within the high and low limits)
1 = Failing (reading has reached or exceeded the high or low limit)
The “1” response message does not tell you which limit (high or low) has been
reached. To determine which limit has failed, you will have to read the measure
-
ment event register (Section 12).
c. CALCulate3:LIMit1:CLEar
CALCulate3:LIMit1:CLEar:AUTO <b>
CALCulate3:LIMit2:CLEar
CALCulate3:LIMit2:CLEar:AUTO <b>
These commands are used to clear the fail (“1”) indications for Limit 1 and Limit
2. If auto clear is enabled for a limit, the fail indication clears when instrument
operation enters the idle state. With auto clear disabled, the fail indication will
remain until it is cleared by the :CLEar command.
Limits and digital outputs programming example
The following command sequence configures the Model 2700 to perform Limit 1 test on a
DCV reading. If the 100mV limit is reached, digital output # 2 will be pulled low. If the
-100mV limit is reached, digital output #1 will be pulled low.
NOTE The following example can be run from the KE2700 Instrument Driver using the
example named “Limits” in
Table H-1 of Appendix H.
*RST ' One-shot measurement mode (DCV).
CALC3:LIM1:UPP 0.1 ' Set HI1 limit to 100mV.
CALC3:LIM1:LOW -0.1 ' Set LO1 limit to -100mV.
CALC3:LIM1:STAT ON ' Enable Limit 1.
CALC3:OUTP:LSEN ALOW ' Set logic sense to active low.
CALC3:OUTP ON ' Enable digital outputs.
READ? ' Trigger and request reading.
CALC3:LIM1:FAIL? ' Request result of limit 1 test.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals