3-56 Basic DMM Operation Model 2700 Multimeter/Switch System User’s Manual
Example #4 — Scan configuration (Model 7700)
The following commands configure scan channels 101, 102, and 121 of a Model 7700
installed in slot 1. When channel 101 is scanned, DCV will be selected. When channel 102
is scanned,
Ω2 will be selected. When channel 121 is scanned, DCI will be selected.
NOTE The following example can be run from the KE2700 Instrument Driver using the
example named “ConigChan” in
Table H-1 of Appendix H.
FUNC 'VOLT',(@101) ' Configure scan channel 101 for DCV.
FUNC 'RES',(@102) ' Configure scan channel 102 for Ω2.
FUNC 'CURR',(@121) ' Configure scan channel 121 for DCI.
NOTE Detailed information on scanning is provided in Section 7.
Measurement queries
NOTE For more information on read commands, see Section 8 (Triggering), Section 13
(SCPI Signal Oriented Measurement Commands), and Appendix D (Signal
Processing Sequence and Data Flow).
:FETCh?
What it does
This command will simply return the latest available reading from an instrument.
Limitations
If the instrument does not have a reading available (indicated by dashes in the display),
sending this command will cause a –230, “Data corrupt or stale” error. This query will not
cause the box to trigger a reading, nor will it “wait” for a result if a reading is in progress.
It is possible to get the same reading over and over using this query. It will continue to give
the same result until one of two things has happened:
A new reading has been triggered
The old reading has been invalidated by changing ranges or by changing function.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals