Model 2700 Multimeter/Switch System User’s Manual Remote Operations 10-3
Separate function setups
A few settings from the front panel are global. That is, the setting on one function also
applies to the other functions. For example, if you set DCV for 3H digits, all the other func
-
tions will also be set to 3H digits. Using remote programming, each function can have its
own unique setup. For example, DCV can be set to 3H digits, ACI can be set to 4H
digits,
and DCI can be set to 5H digits.
Global settings from the front panel that can be set separately using remote programming
include digits, rate, and filter configuration (except on/off, which can be set separately).
NOTE Do not confuse function setups with scan channel setups. For scan channels,
separate settings for digits, rate and filter configuration can be set from either
the front panel or remote programming. See
Section 7 for details on scan
channel setup.
DCV input divider
Using remote programming, you can enable the DCV input divider for the 100mV, 1V, and
10V ranges. When enabled, the input resistance for these DCV ranges are reduced to
10M
Ω. See Section 3 for details on the DCV input divider.
Multiple channel operation
For normal system channel operation, when one measurement channel is closed, the
previous measurement channel opens. With the use of the ROUTe:MULTiple commands,
you gain independent control of all switching module channels, including the relays that
connect the input signal to the DMM. See
Section 2 for details.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD