Model 2700 Multimeter/Switch System User’s Manual Close/Open Switching Module Channels 2-33
Reading relay closure count
To determine the closure count of specific channels, send this query via remote:
ROUTe:CLOSe:COUNt? <clist>
Here, <clist> is the summary of channels. For example, to determine the closure count of
channels 1 and 4 of a module in slot 1, the following query would be sent:
ROUT:CLOS:COUN? (@101,104)
The following query would determine the closure count of slot 1 module channels 1
through 10:
ROUT:CLOS:COUN? (@101:110)
Setting count update interval
Relay closure counts are updated in temporary RAM every time a channel is closed
regardless of how it was closed: by an SCPI command, front panel control, or during a
scan. These counts are permanently written to the EEPROM on the card only at a user-set
time interval (which is initially set to 15 minutes at the factory), or whenever the counts
are queried. Valid intervals (set in integer number of minutes) are between 10 and 1440
minutes (24 hrs).
The lower the interval, the less chance there is of losing relay counts due to power
failures. However, writing to the EEPROM more often may reduce scanning throughput.
The higher the interval, the less scanning throughput is reduced. However, more relay
counts may be lost in the event of a power failure.
NOTE If the Model 2700 is turned off before the updated count is written to EEPROM,
the relay counts will be lost. It is good practice to add the ROUT:CLOS:COUN?
<clist> command at the end of a program to manually update the count.
To set the count update interval, send this command:
ROUTe:CLOSe:COUNt:INTerval <NRf>
where; <NRf> = 10 to 1440 (minutes)
For example, to set the interval to 30 minutes, send this command:
ROUT:CLOS:COUN:INT 30

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals