EasyManua.ls Logo

Keithley 2700 - System:beeper[:State] <B

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2700 Multimeter/Switch System User’s Manual FORMat and Misc SYSTem Commands 14-9
This command is used to simulate front panel key presses. For example, to select the volts
measurement function, send the following command to simulate pressing the “DCV” key:
SYSTem:KEY 2. The key-press codes are also shown in
Figure 14-3.
The queue for the :KEY? query command can only hold one key-press. When :KEY? is
sent and Model 2700 is addressed to talk, the key-press code number for the last key
“pressed” is sent to the computer.
SYSTem:BEEPer[:STATe] <b>
You can disable the beeper for limits and continuity tests. However, when limits or CONT
is again selected, the beeper will automatically enable.
Parameters <b> = 0 Disable beeper
= 1 Enable beeper
Figure 14-3
Key-press codes
1234567816 11
17
18
26 27
19 20
28 29
21 22
30 31
23 24
32 14
15
13
12
F
!
F
Model 2700 Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω 4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT ENTER
DIGITS RATE
RELFILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
OUTPUT
RATIO
ACV
ACI
Ω2 Ω4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS ON/OFFDELAY
HOLD
SAVE SETUP
CONFIG HALT
TYPE
LSYNC
TEST
MONITOR
STEP SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
GPIB

Table of Contents