11-10 Status Structure Model 2700 Multimeter/Switch System User’s Manual
Serial poll programming example
This example is written specifically for the KPCI-488.2 GPIB card and QuickBasic/
VisualBasic with the appropriate IEEE libraries. Other types of cards and/or languages
may have different function calls that are equivalent to the initialize(), transmit(), send(),
srq, and spoll() calls used below.
SRQ when buffer fills with 500, 1000, 1500, and 1750 readings
The following program will store 2000 readings in the buffer. When the buffer fills with
500 readings (quarter full), an SRQ will occur and a message will be displayed on the
computer to indicate that event. An SRQ and message will also occur when the 1000th
(half full), 1500th (three-quarter full), 1750th (buffer notify), and 2000th (full) reading is
stored.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD