EasyManuals Logo

Keithley 2700 User Manual

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
3-58 Basic DMM Operation Model 2700 Multimeter/Switch System User’s Manual
:MEASure[:<function>]?
What it does
This query will reconfigure the instrument to the function specified in the query, set the
trigger source for immediate, set the trigger count to 1, and configure the measurement
parameters to *RST defaults. It will then trigger a single reading, and return the result.
Limitations
This query is much slower than a :READ? or :FETCh? query because it has to reconfigure
the instrument each time it is sent. It will reset the NPLC, autoranging, and averaging to
default settings.
When appropriate
This is an ideal command for taking one-shot measurements if the default settings for a
measurement are appropriate and speed is not a requirement.
[:SENSe[1]]:DATA:FRESh?
What it does
This query is similar to :FETCh? in that it returns the latest reading from the instrument,
but it has the advantage of making sure that it does not return the same reading twice.
Limitations
Like the :FETCh? query, this command does not trigger a reading.
When appropriate
This is a much better choice than the :FETCh? query because it cannot return the same
reading twice. This would be a good query to use when triggering by BUS or
EXTERNAL, because it will wait for a reading to complete if a reading is in progress.
The :CALC:DATA:FRESh? query is similar to the :DATA:FRESh? query, but applies to
readings that have math applied to them (e.g.:MX+B scaling).

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD