10-2 Remote Operations Model 2700 Multimeter/Switch System User’s Manual
Operation enhancements
There are some operations you can do over the IEEE-488 bus and RS-232 interface that
you cannot do from the front panel. The more important ones are summarized below.
Pseudocards
Using remote operation, you can assign a pseudocard to an empty switching module slot.
With a pseudocard installed, the Model 2700 will operate as if the switching module is
installed in the Model 2700. This feature allows you to configure your system without
having the actual switching module installed in the unit. There is a pseudocard for every
Keithley Model 7700 series switching module.
A single SCPI command (SYSTem:PCARdX, where X=1 or 2) is used to install a
pseudocard in an empty switching module slot. Details are provided in
Section 2.
A pseudocard cannot be installed from the front panel. However, once it is installed, you
can take the Model 2700 out of remote and use the front panel. When the instrument is
turned off, the pseudocard will be lost (uninstalled).
Autozero
Autozero is part of the normal measurement process to assure stable, accurate measure-
ments. Autozero can be disabled to increase measurement speed. However, the readings
will eventually become inaccurate over time and temperature changes.
Autozero can only be disabled using remote programming. It cannot be disabled from the
front panel. Autozero is covered in
Section 3.
dB calculation
Using remote programming, you can select the dB calculation for DC or AC voltage. The
dB calculation makes it possible to compress a large range of measurements into a much
smaller scope. See
Section 5 to select and configure the dB calculation.
You cannot select dB from the front panel. However, once it is selected you can take the
Model 2700 out of remote and use the front panel. When the instrument is reset to default
conditions (or turned off), dB will be lost.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals