EasyManuals Logo

Keithley 2700 User Manual

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #307 background imageLoading...
Page #307 background image
10-18 Remote Operations Model 2700 Multimeter/Switch System User’s Manual
RS-232 interface operation
Sending and receiving data
The RS-232 interface transfers data using eight data bits, one stop bit, and no parity. Make
sure the controller you connect to the multimeter also uses these settings.
You can break data transmissions by sending a ^C (decimal 3) or ^X (decimal 18)
character string to the instrument. This clears any pending operation and discards any
pending output.
You can break an RS-232 transmission of buffer readings by pressing LOCAL and then
EXIT. The next command to send buffer data (i.e., TRACe:DATA?) will start at the
beginning, rather than where the transmission was halted.
Baud rate
The baud rate is the rate at which the Model 2700 multimeter and the programming
terminal communicate. Choose one these available rates:
19.2k
9600
4800
2400
1200
600
300
The factory selected baud rate is 4800.
When you choose a baud rate, make sure that the programming terminal that you are
connecting to the Model 2700 can support the baud rate you selected. Both the multimeter
and the other device must be configured for the same baud rate.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD