Model 2700 Multimeter/Switch System User’s Manual Range, Digits, Rate, Bandwidth, and Filter 4-19
Scanning
The moving filter cannot be used when scanning. A scan channel cannot be configured to
use the moving filter. Also, the filter window is not used when scanning.
When a simple scan is configured, the present filter count and state will apply to all
channels in the scan. The window setting is ignored (effectively set to NONE), and if the
moving filter is selected, the filter will not enable when the scan is run.
For the advanced scan, filter state (on or off) and count can be set for each channel. You
cannot set unique filter count, type, and window settings from the advanced scan setup
menu.
For remote programming, the <clist> parameter is used to set filter count and state for
each channel in the scan. You cannot set unique filter type and window settings.
NOTE Details on configuring a scan using filtering in a scan are provided in Section 7.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals