Model 2700 Multimeter/Switch System User’s Manual Buffer 6-3
With buffer auto clear disabled, the only two valid buffer size values are 55000 and
000000. Buffer size 000000 clears the buffer. Entering any other buffer size value resets
the buffer size to 55000.
NOTE If the buffer is empty when the Model 2700 is turned off, buffer auto clear will
enable when it is turned back on.
If the buffer is not empty, the instrument will power up to the last auto clear
setting. Keep in mind that if the instrument powers up with buffer auto clear off,
buffer size is fixed at 55000. You will have to enable auto clear to change the
buffer size.
The auto clear setting (on or off) is not affected by SYSTem:PRESet or *RST
(front panel or remote operation). However, front panel FACTory defaults
enables buffer auto clear.
Enabling/disabling buffer auto clear
1. Press SHIFT and then SETUP.
2. Use the Δ and keys to display the present state of buffer auto clear (BUF
AUTOCLR); Y (yes) or N (no). To retain the present state of buffer auto clear,
press ENTER or EXIT.
NOTE If you change the state of buffer auto clear, the buffer will clear.
3. To change the state of buffer auto clear, press to place the cursor on the present
setting (Y or N).
4. Use the Δ or key to display the desired setting (Y=enabled, N=disabled), and
press ENTER.
NOTE For remote programming, the TRACe:CLEar and TRACe:CLEar:AUTO
commands are used to clear the buffer (
Table 6-1).

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD