Model 2700 Multimeter/Switch System User’s Manual Scanning 7-25
2. Press the Δ or key to display IMM SCAN: N and press ENTER.
a. Press the Δ or key to enable or disable low limit 1 (LLIM1 SCAN:N/Y),
and press ENTER.
b. Press the Δ or key to enable or disable high limit 1 (HLIM1 SCAN:N/Y),
and press ENTER.
c. Press the Δ or key to enable or disable low limit 2 (LLIM2 SCAN:N/Y),
and press ENTER.
d. Press the Δ or key to enable or disable high limit 2 (HLIM2 SCAN:N/Y),
and press ENTER.
3. Finish configuring the scan by performing Step 4 and Step 5 of the Advanced scan
setup procedure,” page 7-16.
NOTE For a remote programmed monitor scan, use the ROUTe:MONitor:POINts
command to specify the number of channels to scan (
Table 7-1).
4. While in the normal measurement state, select and enable the monitor channel as
explained in “
Scan configurationMonitor channel,” page 7-18.
When the reading limit for the monitor channel is reached, the scan will be
triggered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan
finishes, the scan will be triggered to run again. Note that the scan can also be run
by pressing STEP or SCAN.
5. To disable the monitor scan, perform the following steps:
a. To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b. To disable limits, press SHIFT and then ON/OFF. Press Δ or to display
“LIMITS: OFF” and press ENTER.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals