10-12 Remote Operations Model 2700 Multimeter/Switch System User’s Manual
These brackets indicate that IMMediate is implied (optional) and does not have to be used.
Thus, the above command can be sent in one of two ways:
INITiate
or
INITiate:IMMediate
Notice that the optional command is used without the brackets. When using optional
command words in your program, do not include the brackets.
Parameter types — The following are some of the more common parameter types:
<b> Boolean — Used to enable or disable an instrument operation. 0 or OFF
disables the operation, and 1 or ON enables the operation. Example:
SYSTem:LSYNc ON Enable line synchronization
<name> Name parameter — Select a parameter name from a listed group.
Example:
<name> = NEVer
= NEXt
= ALWays
TRACe:FEED:CONTrol NEXt
<NRf> Numeric representation format — This parameter is a number that can be
expressed as an integer (e.g., 8), a real number (e.g., 23.6), or an exponent
(2.3E6). Example:
SYSTem:KEY 11 Press EXIT key from over the bus
<n> Numeric value — A numeric value parameter can consist of an NRf
number or one of the following name parameters; DEFault, MINimum,
MAXimum. When the DEFault parameter is used, the instrument is
programmed to the *RST default value. When the MINimum parameter is
used, the instrument is programmed to the lowest allowable value. When
the MAXimum parameter is used, the instrument is programmed to the
largest allowable value. Examples:
ARM:TIMer 0.1 Sets timer to 100 msec.
ARM:TIMer DEFault Sets timer to 0.1 sec.
ARM:TIMer MINimum Sets timer to 1 msec.
ARM:TIMer MAXimum Sets timer to 99999.99 sec.
<clist> Channel list — Specify one or more channels. Example:
ROUTe:SCAN (@101:110) Scan list; slot 1, channels 1-10
<list> List — Specify one or more numbers for a list. Example:
STATus:QUEue:ENABle (-110:-222) Enable errors -110 through -222

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD