12-8 Common Commands Model 2700 Multimeter/Switch System User’s Manual
I *WAI — Wait-to-Continue Prevent execution of commands until
previous commands are completed
Description
Two types of device commands exist:
Sequential commands – A command whose operations are allowed to finish before the
next command is executed.
Overlapped commands – A command that allows the execution of subsequent
commands while device operations of the Overlapped command are still in progress.
Use the *WAI command to suspend the execution of subsequent commands until the
device operations of all previous Overlapped commands are finished. The *WAI command
is not needed for Sequential commands.
The Model 2700 has three overlapped commands:
:INITiate
:INITiate:CONTinuous ON
• *TRG
NOTE See *OPC, *OPC? and *TRG for more information.
The INITiate commands remove the Model 2700 from the idle state. The device
operations of :INITiate are not considered complete until the Model 2700 returns to idle.
By sending the *WAI command after the INITiate command, all subsequent commands
will not execute until the Model 2700 goes back into idle.
The *TRG command issues a bus trigger that could be used to provide the arm, scan, and
measure events for the Trigger Model. By sending the *WAI command after the *TRG
command, subsequent commands will not executed until the pointer for the Trigger
Model
has finished moving in response to *TRG and has settled at its next state.
Programming example – The following command sequence shows how to use the WAI
command to allow the 2700 to wait for the programmed measurements to be completed
before requesting a reading.
SYST:PRES ‘ Returns 2700 to default setup.
INIT:CONT OFF ‘ Disables continuous initiation.
ABORt ‘ Aborts operation. Places 2700 in idle.
TRIG:COUN 1 ‘ These two commands configure the 2700
SAMP:COUN 30 ‘ to perform 30 measurements.
INIT ‘ Starts measurement process.
*WAI ‘ Sends the WAI command. Program waits for
‘ 2700 to go into idle before executing next command.
DATA? ‘ Requests one reading.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals