EasyManua.ls Logo

Keysight Technologies X Series - Page 113

Keysight Technologies X Series
671 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N9010A EXA Service Guide 113
Instrument Messages
-100 to -199, Command Errors
-128 Numeric data not allowed A legal numeric data element was found, but that is not a valid
element at this position in the command.
-130 Suffix error A problem was found in a suffix (units). The exact problem cannot be
specifically identified.
-131 Invalid suffix There is a syntax problem with the suffix. You need to use the suffix
(units) that are allowed by this command.
-134 Suffix too long The suffix contained more than twelve characters.
-138 Suffix not allowed A suffix was found after a numeric element that does not allow suffixes
(units).
-140 Character data error A problem was found with a character data element. The exact
problem cannot be specifically identified.
-141 Invalid character data Either the character data element contains an invalid character or the
element itself is not valid for this command.
-144 Character data too long The character data element contains more than twelve characters.
-148 Character data not allowed A character data element that you sent is valid, but it is not allowed in
this point in the parsing.
-150 String data error A problem was found with a string data element. The exact problem
cannot be specifically identified.
-151 Invalid string data A string type of data element was expected, but it is invalid for some
reason. For example, an END message was received before the
terminal quote character.
-158 String data not allowed A string data element that you sent is valid, but it is not allowed at this
point in the parsing.
-160 Block data error A problem was found with a block data element. The exact problem
cannot be specifically identified.
-161 Invalid block data A block data element was expected, but it was invalid. For example, an
END message was received before the end length was satisfied.
-168 Block data not allowed A legal block data element was found, but it is not allowed at this point
in the parsing.
-170 Expression error A problem was found with an expression data element. The exact
problem cannot be specifically identified.
-171 Invalid expression An expression data element is not valid. For example, there may be
unmatched parentheses or an illegal character.
-178 Expression data not allowed A legal expression data was found, but it is not allowed at this point in
the parsing.
-180 Macro error A problem was found with a macro element. The exact problem cannot
be specifically identified.
-181 Invalid outside macro
definition
Indicates that a macro parameter placeholder was encountered
outside of a macro definition.