90 Keysight N9010A EXA Service Guide
Instrument Messages
-400 to -499, Query Errors
-400 to -499, Query Errors
Err# Message Verbose/Correction Information
-400 Query Error There was a problem with a query command. The exact problem
cannot be specifically identified.
-410 Query INTERRUPTED Some condition caused an INTERRUPTED query to occur. For
example, a query was followed by DAB or GET before a response was
completely sent.
-420 Query UNTERMINATED Some condition caused an UNTERMINATED query to occur. For
example, the device was addressed to talk and an incomplete program
message was received.
-430 Query DEADLOCKED Some condition caused a DEADLOCKED query to occur. For example,
both the input buffer and the output buffer are full and the analyzer
cannot continue. The analyzer automatically discards output to correct
the deadlock.
-440 Query UNTERMINATED after
indefinite response
A query was received in the same program message after a query
requesting an indefinite response was executed.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals