388 Keysight N9010A EXA Service Guide
Service and Diagnostics Menus
Service Key Descriptions
Band Lock
Provides the ability to tune the analyzer over as large a range as can be
accommodated by the specified harmonic number and 1st LO frequency range.
As a result, this feature can be used to check performance in the frequency
band overlap regions. By definition, multi-band sweeps are not allowed.
When band lock is in effect the Advisory Event “Band Locked: Band <x>” where
“x” is a value between 0 and 6.
When band lock is not in affect the Advisory Event “Band Locked: Band <x>” is
cleared.
Dependencies/Couplings:
1. Band Lock is only available on analyzers with frequency range options
beyond 3.6 GHz
2. The individual bands available for selection also depends upon the
particular frequency range option.
Off or value of -1
Turns Band Lock to Off. Analyzer will tune from band to band
normally.
Band 0 or value of 0
Locks the analyzer to Band 0.
Band 1 or value of 1
Locks the analyzer to Band 1.
Band 2 or value of 2
Locks the analyzer to Band 2.
Band 3 or value of 3
Locks the analyzer to Band 3.
Band 4 or value of 4
Locks the analyzer to Band 4.
Band 5 or value of 5
Locks the analyzer to Band 5.
Band 6 or value of 6
Locks the analyzer to Band 6.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals