536 Keysight N9010A EXA Service Guide
Assembly Replacement Procedures
Power Supply Assembly
Power Supply Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer Case
removal procedure.
2. Remove the rear panel. Refer to the Rear Panel removal procedure.
3. Remove the top brace and reference bracket. Refer to the Top Brace and
Reference Bracket removal procedure.
4. Remove the reference assembly. Refer to the Reference Assembly
removal procedure.
5. Remove the CPU assembly. Refer to the CPU Assembly removal
procedure.
6. Refer to Figure 15-59. Remove the three screws (1) attaching the power
supply (2) to the power supply bracket (3).
7. The power supply assembly can be removed from the chassis by pulling
straight out the back.