524 Keysight N9010A EXA Service Guide
Assembly Replacement Procedures
RF Front End Assembly - Options 532, 544
Replacement
1. Refer to Figure 15-49. Place the RF Front End Assembly into the chassis.
Replace the four screws (2). Torque to 9 inch-pounds.
2. Reattach ribbon cable W17 to the RF Front End Assembly.
3. Reattach the cables W3, W4, W5, W7, W15 (or W36), and W20 to the RF
Front End Assembly (1). Torque the semi-rigid cables to 10 inch-pounds.
4. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
5. Replace the instrument top brace. Refer to the Top Brace and Reference
Bracket replacement procedure.
6. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.