Keysight N9010A EXA Service Guide 117
Instrument Messages
Condition Errors 1 to 99, Calibration
Condition Errors 36 to 64, Calibration Needed or Failed
Errors with instrument internal alignment routines.
An event with the error number shown in the table means the condition has
been detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 42 indicates an RF Alignment Failure has been detected,
error 1042 indicates that failure has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to
SCPI. Nothing in this column means nothing is put up; status bit only.
Err# Message Error or
Warning
Verbose/Correction Information
36 DCFM Zero Failure E
38 IQ Calibration Failure E
40 TG Alignment Failure E
42 RF Alignment Failure E
44 IF Alignment Failure E
46 LO Alignment Failure E
48 ADC Alignment Failure E
50 FM Demod Alignment Failure E
52 Extended Align Needed Sum E
54 Extended Align Failure Sum E
56 Tracking Peak Needed W
58 Align Skipped Sum E
60 Align RF Now Needed E
62 Corrections Off W
64 Align All Now Needed E

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals