Keysight N9010A EXA Service Guide 119
Instrument Messages
Condition Errors 1 to 99, Calibration
Condition Errors 67 to 95, Calibration Failure (Extended)
Errors with instrument internal alignment routines failing.
An event with the error number shown in the table means the condition has
been detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 67 indicates that the RF Alignment not being done during
alignments has been detected, error 1067 indicates that failure has been
cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to
SCPI. Nothing in this column means nothing is put up; status bit only.
Err# Message Error or
Warning
Verbose/Correction Information
67 Align 9kHz to 30MHz failed W
69 Align 30MHz to 1GHz failed W
71 Characterize Preselector
failure
W The preselector characterization routine failed.
73 unused bit4 is set
75 unused bit4 is set
77 unused bit5 is set
79 unused bit6 is set
81 unused bit7 is set
83 unused bit8 is set
85 unused bit9 is set
87 unused bit10 is set
89 unused bit11 is set
91 unused bit12 is set
93 unused bit13 is set
95 unused bit14 is set