294 Keysight N9010A EXA Service Guide
Analog/Digital IF Troubleshooting
40 MHz BW IF Section
A3 Digital IF Troubleshooting
Verifying the 22.5 MHz Input Power
1. Perform an instrument shutdown.
2. Remove the cover of the analyzer. Refer to Chapter 15, “Assembly
Replacement Procedures” in this service guide.
3. Turn the instrument over so that the bottom side of the analyzer is facing
up.
4. Turn on the analyzer and wait for the instrument to complete the boot up
process.
5. Press System, Alignments, Auto Align, Off
6. Press Input/Output, RF Calibrator, 50 MHz
7. Verify the 50 MHz signal is at 25 dBm by pressing FREQ, 50 MHz, SPAN,
1 MHz, Peak Search on the analyzer. The marker readout should be
50 MHz at 25 dBm ± 3 dB. If this reference signal is measuring
incorrectly, see Chapter 4, “RF Section Troubleshooting (RF/Microwave
Analyzers)” in this service guide.
8. To continue verifying press SPAN, Zero Span. Verify the input attenuator
on the analyzer is set to 10 dB. Look near the top of the display near the
center and verify Atten: 10 dB is visible. If needed change the input
attenuator by pressing AMPTD, Attenuation, 10 dB on the analyzer.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals