Keysight N9010A EXA Service Guide 477
Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
Low Band Switch -
Standard Instruments with Serial Number < MY/SG50510001
Removal
1. Refer to Figure 15-10. Remove the ribbon cable W18.
2. Remove the semi-rigid cables W3, W8, and W9 using the 5/16 inch
wrench.
3. Remove the two screws (2) using the T-10 driver. The low band switch
(1) can now be removed from the chassis.
W8 will not be present in instruments with Option 503.