Keysight N9010A EXA Service Guide 529
Assembly Replacement Procedures
L.O. Synthesizer Assembly
Replacement
1. Refer to Figure 15-52. Install the L.O. Synthesizer assembly into slot 5 in
the chassis and press down to plug it into the motherboard.
2. Reattach the cables W4 and W6 to the L.O. Synthesizer assembly (1).
Torque to 10 inch-pounds.
3. Replace the instrument top brace. Refer to the Top Brace and Reference
Bracket replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.