Keysight N9010A EXA Service Guide 41
Overview
After an Instrument Repair
After an Instrument Repair
If any instrument assemblies have been repaired or replaced, perform the
related adjustments and performance verification tests. These tests are done
using the N7814A Keysight X-Series Signal Analyzer Calibration Application
Software. Refer to Chapter 16, “Post-Repair Procedures” for a list of
post-repair adjustments and performance tests based on which hardware has
been changed.
Information regarding the N7814A Keysight X-Series Signal Analyzer
Calibration Application Software can be found at
http://www.keysight.com/find/calibrationsoftware