28 Keysight N9010A EXA Service Guide
Overview
Service Equipment You Will Need
Required Test Equipment List
The following table identifies the equipment recommended for
troubleshooting, adjusting, and verifying the performance of the instrument.
Only the recommended and alternate equipment is compatible with the
performance verification testing. Some tests can use various models of a
particular equipment type. The “Recommended Keysight Model” is the
preferred equipment. However, the “Alternative Keysight Model” is an
acceptable substitute.
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Model
a
Alternative
Model
Use
b
Signal Sources
Microwave Signal
Generator #1
Frequency: 10 MHz to 50 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −30 dBc
Amplitude range: −40 dBm to +13
Amplitude resolution: 0.02
VSWR:
< 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
E8257D
(Option 567,
1EA/1EU, 1E1)
(Option 007,
Analog Ramp
Sweep,
required for
Frequency
Response
Adjustments)
For Adjustments:
83630A/B/L
(Option 001, 008),
83640A/B/L
(Option 001, 008),
83650A/B/L
(Option 001, 008)
PSG
c
For Performance
Tests:
83630A/B
(Option 001, 008),
83640A/B
(Option 001, 008),
83650B
(Option 001, 008)
PSG
c
A, P, T
Microwave Signal
Generator #2
(only required for Third
Order Intermodulation
and Gain Compression)
Frequency: 10 MHz to 26.5 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −30 dBc
Amplitude range: −20 dBm to +13
Amplitude resolution: 0.02
VSWR:
< 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
E8257D
(Option
1EA/1EU)
83630A/B
(Option 008),
83640A/B
(Option 008),
83650A/B
(Option 008)
PSG
c
P