Keysight N9010A EXA Service Guide 121
Instrument Messages
Condition Errors 101 to 199, Measurement Integrity
Condition Errors 101 to 199, Measurement Integrity
Errors with making measurements: triggering, over range, bad
acquisition/data, bad settings.
An event with the error number shown in the table means the condition has
been detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 129 indicates a Meas Uncal condition has been detected,
error 1129 indicates that failure has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to
SCPI. Nothing in this column means nothing is put up; status bit only.
Err# Message Error or
Warning
Verbose/Correction Information
133 Signal Summary E
135 No Result E
135 No Result; Meas invalid
with I/Q inputs
E The current measurement does not support I/Q input;
switch to the RF or another input or select a different
measurement
135 No Result;Turn on MCE E To calculate Timing and Phase results in the Code Domain
Power view of Mod Accuracy, the "Multi Channel Estimator"
must be set to ON. Otherwise these results are invalid.
137 (unused)
139 Uncalibrated Summary E
141 Input Overload W
141 Input Overload;ADC
over range
W The signal at the input to the IF section is too high. You
should increase the attenuation or lower the signal level.
141 Input Overload;I/Q ADC
over range
W The I or Q input exceeds the ADC upper limit.
141 Input Overload;I/Q
Voltage over range
W The input voltage on the I or Q channel exceeds the
channel limit. In differential mode the over voltage may
occur without causing an ADC overload, for example, if I is
at +5.01 V and I-bar is at +5.0 the ADC will be in range but
both I and I-bar will exceed the voltage limit.
143 (unused)
145 (unused)