594 Keysight N9010A EXA Service Guide
Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Table 16-1 lists the adjustments and performance verification tests needed
after an assembly replacement.
After an assembly is replaced, find the assembly that has been replaced in the
left-hand column, and then perform the recommended adjustment and/or
performance verification test. Making the recommended adjustments and/or
performance verification tests doesn’t guarantee all published specifications
are being met. Only a full instrument calibration will do this.
The following procedures are also included in the Additional Tasks” on
page 598.
“Calibration Data Backup and Restore”
“BIOS Settings Verification
“End User License Agreement (EULA) Acceptance (Software versions
A.06.08)”
“End User License Agreement (EULA) Acceptance (Software versions
A.07.00)”
“FPGA Synchronization
“Programming Model and Serial Numbers”
Refer to the analyzer user’s guide for information on instrument warm-up before performing any
of the procedures listed in this chapter.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals