Keysight N9010A EXA Service Guide 539
Assembly Replacement Procedures
CPU Assembly
CPU Assembly
Removal
Depending on the vintage of your instrument, the CPU assembly will look
either like Figure 15-60 or Figure 15-61.
Figure 15-60 Current CPU Assembly - Rear Panel View
Figure 15-61 Old CPU Assembly - Rear Panel View
1. Refer to Figure 15-62 or Figure 15-63. Remove the six screws (1)
attaching the CPU assembly to the chassis.
2. The CPU assembly can be removed from the chassis by pulling straight out
the back. Use the two ejectors to pull the CPU assembly out from the
chassis.
If the A4BT1 CPU board battery was changed, refer to Chapter 16, “Post-Repair
Procedures.” for setup instructions.