Keysight N9010A EXA Service Guide 541
Assembly Replacement Procedures
CPU Assembly
Replacement
1. Slide the CPU assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the midplane assembly.
Secure the board with the ejectors.
2. Refer to Figure 15-62 or Figure 15-63. Replace the six screws (1) that
attach the CPU assembly to the chassis. Torque to 9 inch-pounds.