Keysight N9010A EXA Service Guide 559
Assembly Replacement Procedures
Analog I.F. (AIF)/Digital I.F. (DIF) Assembly
Replacement
1. Slide the AIF/DIF assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the midplane assembly.
2. Refer to Figure 15-76. Replace cables W13, W14, and W15 or W37 to the
AIF assembly.
3. Refer to Figure 15-74 and Figure 15-75. Replace the appropriate cables to
the bottom of the DIF assembly.
4. Replace the rear panel. Refer to the Rear Panel replacement procedure.
5. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.